3DATX Corporation

Flexible iPEMS: Design The Best Solution For YOU

The 3DATX parSYNC® iPEMS (integrated Portable Emissions Measurement System) provides gaseous and particulate measurements with a proprietary cartridge system. The “hot-swap” capability delivers emissions acquisition of 4-gas measurement, NOx measurement and nanoparticle measurement.

The 3DATX patented multi-plex basic particle sensor system provides the ability to capture multiple, dissimilar “images” of particles using ionization, scattering, and opacity sensors in addition to outputs for particle number (PN) and particulate mass (PM).

FLEX Cartridge Configuration

The parSYNC® FLEX is designed with 3 sensor cartridge ports, allowing you to customize each unit for your specific testing needs:

Further customize your parSYNC® FLEX to allow for real-time acquisition of:

GasesNon-Dispersive Infrared (NDIR)Individual Electro-Chemical Cells
CO2COHCNONO2O2
Range0-20%0-15%0-4000 ppm0-5000 ppm0-300 ppm0-100%
T90 Time< 3.5 s< 3.5 s< 3.5 s< 5 s< 35 s< 6 s
Resolution0.01%0.01%1 ppm1-2 ppm0.1 ppm0.01 ppm
AccuracyAbs.±0.3%±0.02%±8 ppm±15 ppm±5 ppm±0.1%
Rel.±3%±3%±3%±2%±2%±2%
RepeatabilityAbs.±0.1%±0.02%±6 ppm±0.1%±0.1%±0.1%
Rel.±2%±2%2%2%±0.1%±0.1%
Particulates
DetailsData
Particle Size Range10 to 10,000 nm = 0.01 to 10 µm
Ionization SensorUltra-Fine: 10 to 250 nm / peak @ < 80 nm
Opacity SensorMedium: 80 to 7,000 nm / peak @ ~800 nm
Scattering SensorCoarse: 250 to 10,000 nm / peak @ ~2,500 nm

parSYNC® FLEX-PNC iPEMS

The 3DATX parSYNC® FLEX-PNC iPEMS provides gaseous AND particle number measurements. The gaseous side utilizes a “hot-swap” cartridge system for traditional 4-gas measurement (CO2, CO, HC, O2) plus NOx (NO, NO2). The particulate matter side utilizes NMi, METAS, and PTB certified particle number concentration (PNC) sensor based on the extended diffusion charging principle.

PNC Sensor Specifications
DetailsData
Range5,000 - 5,000,000 #/cm3
Resolution1,000 #/cm3
Accuracy25,000 #/cm3 absolute or 25% relative
Counting Efficiency23 nm ±5%: 20-60%
50 nm ±5%: 60-130%
80 nm ±5%: 70-130%

Features and Benefits

CUBE™ FLEX (Conditioning Unit for Batch Emissions)

Support unit connected to the parSYNC® for sample conditioning and extended power supply:

3DATX Corporation

501 John James Audubon, Suite 200
Buffalo NY 14228
USA

Phone: +1 844.30.3DATX (3289)
Web: 3datx.com
Email: info@3datx.com